Megger EVCC300 – EV Charger Checker
Revolutionize your maintenance, repair, and manufacturing check routines with the all-new Megger EVCC300 Charger Checker. Designed specifically for Electric Vehicle chargers, this advanced tool offers a streamlined and user-friendly solution for checking the safety and operation of mode 2 and single-phase mode 3 chargers.
Key for those seeking to perform fault-checks post-manufacture, post-repair, or as part of a regular maintenance schedule, the EVCC300 brings convenience and reliability to your fingertips.
Unmatched Features for Comprehensive Checks
The EVCC300 delivers on four pivotal safety checks, ensuring a hazard-free charging experience. These include a PE touch pad for grounding and voltage checks, a PE resistance test for exposed metal work grounding, an essential touch voltage test, and an RCD/GFCI protection check that validates the device tripping speed.
Moreover, our unique product helps to identify potential nuisance charger supply tripping which could disrupt your EV charging experience.
Delivering Operational Excellence
The EVCC300 stands apart with its four key EV charger operation checks. These include proximity detection operation check, control pilot operation check conforming to IEC61851-1 requirements, control pilot signal check, and EV charger output check.
Key Features At-a-Glance
- Compatible with mode 2 and single-phase mode 3 chargers
- Compatible with SAE J1772 Type 1 and Type 2 connectors
- Effective in performing four essential EV charger safety checks
- Ensures four crucial EV charger operation checks
- Assists in identifying if nuisance tripping or RCD/GFCI is likely to occur
- Reads control pilot state response from EV Charger to IEC61852-1 standards
Invest in the Megger EVCC300 Charger Checker for a seamless, safe, and efficient EV charging experience. Elevate your EV charger maintenance and safety checks to the next level today!
SPECIFICATIONS
SPECIFICATIONS
EVSE STANDARD IEC 61851-1:2017
EVSE SUPPLY OPTIONS (set in settings)
- 230 V single phase (EU / UK)
- 120 V single phase (US)
- 208 V / 240 V two phase (US)
SUPPLY MEASUREMENT
Voltage range: 5 V to 300 V
Voltage resolution: 1 V
Voltage accuracy: ± 5%, ± 2 digits
Voltage measurement: L to N, L to PE and N to PE
Reverse polarity indication: L to N, for single phase option only
Frequency range 45 Hz to 65 Hz
Frequency resolution: 1 Hz Frequency accuracy ± 1 digit
EVSE INTERFACE (EV charger function) tests:
1. Proximity detection (PP) states
Read only
2. Control pilot (CP)
No proximity
Proximity detected, unlatched Proximity detected, latched
Set and read states: State A – Disconnected
State B – Connected State C – Charging
State D – Charging without ventilation
State E – CP short to PE
Read only state State F – EVSE charger fault condition
Maximum voltage ± 14 V
Frequency range 940 Hz to 1040 Hz
Duty cycle range 8% to 97%
Maximum charging current indication according to IEC 61851-1:2017, tables A.7 and A.8
RESISTANCE OF EARTH PROTECTIVE CONDUCTOR (RPE)
Test current: 200 mA
Resistance range: 0 to 10 Ω Resistance resolution: 0.01 Ω Resistance accuracy: ± 5%, ± 2 digits
RCD/GFCI TESTS:
Trip time accuracy: ± 1%, ± 1 ms
Trip current accuracy: ± 3% (apply to all RCD/GFCI tests)
PERSONAL PROTECTION TESTS, 230 V SINGLE PHASE:
- RCD 30 mA AC TEST
Test current: 31.5 mA AC
(5% above nominal trip current) Test current accuracy: ±1.0 mA
Max. test time: 300 ms
Polarity selection: 0° and 180° of an input sine wave
- RCD 6 mA DC TEST
Test current ramp up at rate of 6 mA in 2.5 s,
then held at 6 mA DC Test current accuracy: ±0.5 mA
Ramp polarity: positive and negative ramp
Max. test time: 12.5 s
Polarity selection: 0° and 180° of an input sine wave
PERSONAL PROTECTION TESTS, 120 V SINGLE PHASE / 240 V TWO PHASE
- GFCI/CCID 5 mA AC test
Test current: 6 mA AC Test current accuracy: ±0.5 mA Max. test time: 5.59 s
Polarity selection: 0° and 180° of an input sine wave
- GFCI/CCID 20 mA AC test
Test current: 21 mA AC
(5% above nominal trip current) Test current accuracy: ±0.5 mA
Max. test time: 5.59 s
Polarity selection: 0° and 180° of an input sine wave
NUISANCE TRIPPING TESTS, 230 V SINGLE PHASE
- RCD AC ramp test
Test current AC current ramp up to 30 mA in 2 mA steps
Test current accuracy: ±0.5 mA Step time: 300 ms
Max. test time: 4.5 s
- RCD DC ramp test
Test current ramp up at rate of 6 mA in 2.5 s,
then held at 3 mA DC.
Max. test time: 11.25 s
NUISANCE TRIPPING TESTS, 120 V SINGLE PHASE / 240 V TWO PHASE
- GFCI/CCID 5 mA AC test
Test current AC current ramp up to 6 mA in 0.5 mA steps
Test current accuracy: ±0.5 mA Step time: 100 ms
Max. test time: 1.2 s
- GFCI/CCID 20 mA AC test
Test current AC current ramp up to 20 mA in 1 mA steps
Test current accuracy: ±0.5 mA Step time: 100 ms
Max. test time: 2 s
TOUCH VOLTAGE TEST
Options: 25 V or 50 V limit
Test current typically 1/3 of RCD nominal test current
TOUCH CONTACT PAD TEST
Dangerous voltage indication at PE conductor
POWER SUPPLY
Battery: Four AA cells
CONNECTIONS
Type 2 male – main test connection
4 mm shrouded socket on base of handle – RPE test return connection
LANGUAGES
User Interface: English, French, German and Spanish
User guide: English, French, German and Spanish
DIMENSIONS AND WEIGHTS
Dimensions are (L x H x W)
24 x 18 x 8.6 cm.
Weight (without batteries)
583 g
SAFETY
Standards IEC 61010-1:2010
Safety category CAT II 300 V
ENVIRONMENT
Operational temperature: 0 ºC to 40 ºC Storage temperature -10 ºC to 70 ºC
Operating humidity 90% R.H. at +40 °C max Maximum altitude: 2000 m
Dust and moisture ingress: IP40 to IEC60529 in use